ASCII Protocol 15
2.5 Examples
2.6 Error Messages
:CALSniff calibration factor sniff 521 R/S
:CALVac calibration factor vacuum 520 R/S
*SERVICE
:READBuffer Read service buffer
1300 ..
1310
R
*STARTFLA
SH
Flash-Update starten 2619 S
*RST :FACTORY Sets all parameters to factory default 1161 S
:CALHistory Clears calibration history 1161 S
:ERRORHistory Clears error history 1161 S
Relates to
LD cmd. no.
Read /
Set
Command answer
*stat? (CR) MEAS (CR) mode
*status? (CR) MEAS (CR) mode
*read? (CR) 2.876E-7 (CR)
leak rate according to
programmed unit
*read:pa*m3/s? (CR) 2.876E-6 (CR) leak rate in a different unit
*start (CR) OK (CR) start measurement
*conf:trig1? (CR) 1.0E-9 (CR) retrieve trigger 1
*conf:trig1 2.0E-9 (CR) OK (CR) set trigger 1
Message Meaning
OK command completed
E01 wrong command start (no „*“)
E02 illegal blank
E03 command word 1 illegal
E04 command word 2 illegal
E05 command word 3 illegal
E06 control by RS232 not enabled
E07 argument faulty
E08 no data available
E09 error buffer overflow
E10 command invalid
E11 query not allowed
E12 only query allowed
E13 not yet implemented