OPERATING MANUALCOMPOSERGas Concentration ControllerIPN 074-289®

Strany 3

4 - 4 IPN 074-289L Composer Operating Manual4.2.4 How To Revert To Factory Default SettingsThe factory set Reference Zero is re-established by chang

Strany 4

4 - 5IPN 074-289LComposer Operating Manual4.3 How To Use The InstrumentThe following sections explain how to start, stop, and shut down the instrum

Strany 5 - 09/24/97

4 - 6 IPN 074-289L Composer Operating Manual4.3.2.2 Bulk Flow Control ConfigurationAll instrument function is controlled externally through Input #4

Strany 6

4 - 7IPN 074-289LComposer Operating Manual4.3.4.2.2 Bulk Flow Control ConfigurationRefer to section 3.1.13 on page 3-11 for Bulk Flow Control speci

Strany 7 - Registration Card

4 - 8 IPN 074-289L Composer Operating ManualSTEADY . . . . . . . . . . . . . . . . . . . . A green LED that is illuminated when the measured concent

Strany 8 - BUSINESS REPLY MAIL

4 - 9IPN 074-289LComposer Operating Manual4.3.5.2 Optional - Composer LCD Display DescriptionThe LCD display’s layout is shown in Figure 4-3. The d

Strany 9 - Warranty

4 - 10 IPN 074-289L Composer Operating Manual4.3.5.2.1 Numerical Display and Units Indicators-.8.8.8.8.8.8.8 . . . . . . . . . . . . . . . . Numeri

Strany 10

4 - 11IPN 074-289LComposer Operating Manual4.3.5.2.2 State IndicatorsSTOP . . . . . . . . . . . . . . . . . . . . . . Illuminates when the instrum

Strany 11 - Table Of Contents

4 - 12 IPN 074-289L Composer Operating Manual4.3.5.2.3 Instrument Zero IndicatorsFACTORY ZERO . . . . . . . . . . . . . . Illuminates when the instr

Strany 12 - How the Instrument Works

4 - 13IPN 074-289LComposer Operating ManualÚ . . . . . . . . . . . . . . . . . . . . . . . . . Illuminates when the concentration meets the condit

Strany 13 - Applications Guide

TOC - 1IPN 074-289LComposer Operating ManualTable Of ContentsChapter 1Getting Started1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . .

Strany 14 - Operation

4 - 14 IPN 074-289L Composer Operating Manual4.3.5.2.6 Film Number IndicatorPROG # . . . . . . . . . . . . . . . . . . . . . Always illuminated. Ind

Strany 15

4 - 15IPN 074-289LComposer Operating Manual4.3.5.2.7 Program Parameter IndicatorsTable 4-1 is also useful for the range and limits set by the Setup

Strany 16 - TOC - 6

4 - 16 IPN 074-289L Composer Operating ManualCONCENTRATION TARGET . . . . . The desired concentration of precursor in terms of mole %. Also used to m

Strany 17

4 - 17IPN 074-289LComposer Operating ManualMFC RANGE DIL . . . . . . . . . . . . . The parameter representing the full flow range of the mass flow c

Strany 18

4 - 18 IPN 074-289L Composer Operating ManualLCD display for the specific location’s lighting and instrument position, see section 5.8, LCD Contrast

Strany 19 - Getting Started

4 - 19IPN 074-289LComposer Operating Manual4.3.5.3.1 Function and Numeric KeysThe data entry and process control portion occupies the upper right p

Strany 20 - 1.2 Using this Manual

4 - 20 IPN 074-289L Composer Operating ManualNOTE: Holding the C key during a "power on" will unlock parameters if they have been locked by

Strany 21

4 - 21IPN 074-289LComposer Operating ManualS. . . . . . . . . . . . . . . . . . . . . . . . . . Pressing this key will temporarily convert the nume

Strany 22 - 1.4.1 Application Support

4 - 22 IPN 074-289L Composer Operating Manual4.3.6 Description of Rear PanelThe rear panel of the Composer is shown in Figure 4-5. The rear panel is

Strany 23

4 - 23IPN 074-289LComposer Operating ManualThe eight inputs are active when the specific input’s terminal is pulled to ground (common) (<0.8 V) t

Strany 24 - 1.6 Physical Requirements

TOC - 2 IPN 074-289L Composer Operating Manual1.10.4.1 When Used as a Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Strany 25

4 - 24 IPN 074-289L Composer Operating Manual4.3.6.2 SENSOR ConnectorA DB-15 connector used for connecting the Controller to the Transducer. There a

Strany 26 - 1.7.2 Rated Input

4 - 25IPN 074-289LComposer Operating Manual4.3.6.3 POWER ConnectorA DB-9 connector used for connection to a 24 volt power supply. Refer to section

Strany 27 - 1.10 Installation

4 - 26 IPN 074-289L Composer Operating Manual4.3.6.6 CFGR2The second eight contact binary encoded switch block that is used for activating various i

Strany 28 - 1.10.1 Physical Installation

4 - 27IPN 074-289LComposer Operating Manual4.3.6.7 RS-232CSee Chapter 7 for information regarding RS-232C.4.3.6.8 CONCENTRATION OUT ConnectorA BNC

Strany 29 - 1 - 11

4 - 28 IPN 074-289L Composer Operating Manualand is updated every 200 milliseconds. It is intended for driving a chart recorder or being read by a sy

Strany 30

4 - 29IPN 074-289LComposer Operating Manual4.3.6.11 Analog I/O OptionAn optional Analog I/O card is available to support the Bulk Flow Control conf

Strany 31 - 1 - 13

4 - 30 IPN 074-289L Composer Operating ManualThis page is intentionally blank.

Strany 32 - Requirements

5 - 1IPN 074-289LComposer Operating ManualChapter 5Maintenance5.1 Safety ConsiderationsWARNINGMany of the complex chemicals that will be flowed thr

Strany 33 - 1 - 15

5 - 2 IPN 074-289L Composer Operating Manual5.5 Inspecting the TransducerIf the Transducer is disassembled, all components should be visually inspec

Strany 34 - 1 - 16

5 - 3IPN 074-289LComposer Operating Manual5 Close the Leak Test Ports with a copper gasket following the manufacturer’s suggested tightening procedu

Strany 35 - 1 - 17

TOC - 3IPN 074-289LComposer Operating ManualChapter 3Applications Guide3.1 Advice and Tips. . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strany 36 - 1 - 18

5 - 4 IPN 074-289L Composer Operating Manual3 Probe with a very small amount of helium into the appropriate end of the transducer. Small openings are

Strany 37 - 1 - 19

5 - 5IPN 074-289LComposer Operating ManualFigure 5-1 Correctly operatingFigure 5-2 Possible diaphragm damageFigure 5-3 Possible diaphragm damage

Strany 38

5 - 6 IPN 074-289L Composer Operating Manual5.8 LCD Contrast Adjustment —(LCD Display Option Only)The LCD contrast is optimized for "above the

Strany 39 - 1.10.5 Sensor Installation

6 - 1IPN 074-289LComposer Operating ManualChapter 6Diagnosing Problems6.1 IntroductionThere are many built in checks and diagnostics in the firmwar

Strany 40 - 1 - 22

6 - 2 IPN 074-289L Composer Operating Manual6.2.1 Category 1; Temperature Related*Temperature LED flashes rapidly when the cell’s temperature is mor

Strany 41 - 1 - 23

6 - 3IPN 074-289LComposer Operating Manual6.2.2 Category 2; Acoustic Generator RelatedNOTE: PRT stand for Platinum Resistance Thermometer, C stands

Strany 42 - 1 - 24

6 - 4 IPN 074-289L Composer Operating Manual6.2.4 Category 4; Installation RelatedTable 6-4 Category 4; Installation RelatedFailure Symptom Visibil

Strany 43

6 - 5IPN 074-289LComposer Operating Manual6.2.5 Category 5; User RelatedTable 6-5 Category 5; User RelatedFailure Symptom Visibility Effect On Acc

Strany 44

6 - 6 IPN 074-289L Composer Operating Manual6.2.6 Category 6; Miscellaneous6.3 How To Revert To Factory Default ValuesRefer to section 4.2.4, How T

Strany 45

6 - 7IPN 074-289LComposer Operating Manual6.5 User DiagnosticsVarious failure modes can be identified by (1) carefully watching the WARNING LED loc

Strany 46 - vs. Instrument’s Frequency

TOC - 4 IPN 074-289L Composer Operating Manual3.2.1.3 Typical Parameter Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2

Strany 47

6 - 8 IPN 074-289L Composer Operating ManualHeater failure/FatalThe temperature has been dropping for 10 seconds but the heaters are on full.Rapid Bl

Strany 48 - 2.3 Control Loop Primer

6 - 9IPN 074-289LComposer Operating Manual6.5.1 How To Start User DiagnosticsThe User Diagnostics built into the controller operate automatically.

Strany 49 - (control signal is increased)

6 - 10 IPN 074-289L Composer Operating Manual6.6 Error DetectionSpecific errors can be detected in one of the following ways:1 Observing the blink p

Strany 50 - IAE e t() td

7 - 1IPN 074-289LComposer Operating ManualChapter 7External Communications7.1 IntroductionThe Composer supports RS-232C with Checksum.7.2 Communic

Strany 51 - 2.4 Fuzzy Logic

7 - 2 IPN 074-289L Composer Operating Manual7.3 Communication SettingsSerial communications baud rates between 4800 and 57600 are supported and sele

Strany 52 - 2 - 10

7 - 3IPN 074-289LComposer Operating Manual7.4.1 Structure of a Command Sent to ComposerLength Command [Subcommand] [Program Number] [Data] Checksum

Strany 53 - Fuzzy Logic Control

7 - 4 IPN 074-289L Composer Operating Manual7.4.2 Structure of a Response Sent from ComposerLength Status Data ChecksumLength is one byte, and does

Strany 54 - 2.5 Instrument Overview

7 - 5IPN 074-289LComposer Operating Manual7.4.3 List of Commands with their Subcommands, Data, and Response Data.Command Sent Data Response DataSub

Strany 55 - 2.6.1 Transducer

7 - 6 IPN 074-289L Composer Operating ManualU(pdate)Data - value of the parameter specified by the subcommand.No response data unless there is a comm

Strany 56 - 2 - 14

7 - 7IPN 074-289LComposer Operating ManualS(tatus) 00 Current Data float float float float byte byte(frequency amplitude concentration/phase tem

Strany 57 - 2.6.2 Controller

TOC - 5IPN 074-289LComposer Operating Manual4.3.5.2 Optional - Composer LCD Display Description . . . . . . . . . . . . . . . . . . .4-94.3.5.2.1 N

Strany 58 - 2 - 16

7 - 8 IPN 074-289L Composer Operating ManualCodes for S and R 990 Frequency Float1 Amplitude Float2 Temperature Float3 Concentration Floa

Strany 59 - 2 - 17

7 - 9IPN 074-289LComposer Operating ManualR(emote)No data sentNo Response data unless there is a communication error.Performs an action depending on

Strany 60 - 2.6.5 Setup Tool Software

7 - 10 IPN 074-289L Composer Operating Manual7.4.4 Error Codes7.4.4.1 Communication ErrorsIf there was an error with the command send to the Compos

Strany 61 - 2 - 19

7 - 11IPN 074-289LComposer Operating Manual7.4.4.2.1 System and Hardware Errors0x8000 . . . . . . . . . . . . . . . . . . . . . Non Volatile Memor

Strany 62 - 2 - 20

7 - 12 IPN 074-289L Composer Operating Manual7.4.5 ExamplesNOTE: Commands and responses are all using hexadecimal representation.Example 1Send Hello

Strany 63

8 - 1IPN 074-289LComposer Operating ManualChapter 8Software Guide8.1 IntroductionThe two software applications available for interfacing the Compos

Strany 64

8 - 2 IPN 074-289L Composer Operating Manual8.2.1 Computer System Requirements8.2.2 Software InstallationEither the Composer Monitor or the Setup T

Strany 65

8 - 3IPN 074-289LComposer Operating Manual8.2.3 First Time UseAfter you have connected Composer to a free COM port on your computer and applied pow

Strany 66

8 - 4 IPN 074-289L Composer Operating Manual8.2.3.3 Set Up Process ParametersOnce you have set up the communications parameters,1 Go to the Setup &g

Strany 67 - Mode Be Used?

8 - 5IPN 074-289LComposer Operating Manual8.2.3.4 Using the ApplicationsBoth applications are primarily used in the Amplitude Track mode for real t

Strany 68

TOC - 6 IPN 074-289L Composer Operating ManualChapter 6Diagnosing Problems6.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strany 69

8 - 6 IPN 074-289L Composer Operating Manual8.3 Software ReferenceThis section describes the function of every button in the Composer Monitor and Se

Strany 70 - Date (1997)

8 - 7IPN 074-289LComposer Operating ManualSetup Communications Brings up a dialog box for setting up communications with Composer. See section 8.3.4

Strany 71 - Ar 6.1 4.1

8 - 8 IPN 074-289L Composer Operating ManualActions Resume CollectionInitiates or resumes data collection in current mode of operation.End Collection

Strany 72 - 3 - 10

8 - 9IPN 074-289LComposer Operating ManualCalibration Set User Zero Sets instrument zero. Should be done only when pure carrier gas is flowing at st

Strany 73 - 3 - 11

8 - 10 IPN 074-289L Composer Operating Manual8.3.2 ToolbarFigure 8-2 ToolbarThe Main Toolbar provides easy access to the two measurement modes and

Strany 74 - 3 - 12

8 - 11IPN 074-289LComposer Operating ManualAboutClicking this button shows the version number of software as well the version of firmware in the Com

Strany 75 - 3 - 13

8 - 12 IPN 074-289L Composer Operating Manual8.3.3 Ribbon BarFigure 8-3 Ribbon BarThe Ribbon bar provides access to often used functions. If the Ri

Strany 76 - 3 - 14

8 - 13IPN 074-289LComposer Operating Manual8.3.4 Setup Communications ParametersThese parameters are necessary to set up communications between the

Strany 77 - 3 - 15

8 - 14 IPN 074-289L Composer Operating Manual8.3.5 Setup Process ParametersThese parameters are necessary to set the sensor’s temperature, provide g

Strany 78 - 3 - 16

8 - 15IPN 074-289LComposer Operating Manual8.3.5.1 IndexProgram Number. . . . . . . . . . . . . The Composer instrument can store up to nine sets

Strany 79 - 3 - 17

TOC - 7IPN 074-289LComposer Operating ManualChapter 8Software Guide8.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strany 80 - 3 - 18

8 - 16 IPN 074-289L Composer Operating Manual8.3.5.3 Dilution MFC (sccm)Full Scale: . . . . . . . . . . . . . . . . . . A numeric data field repres

Strany 81 - 3.1.14 Analog I/O Option —

8 - 17IPN 074-289LComposer Operating ManualDead Time: (s) . . . . . . . . . . . . . . . A numeric data field representing the process dead time. Th

Strany 82 - 3 - 20

8 - 18 IPN 074-289L Composer Operating Manual8.3.5.8 Precursor GasName: . . . . . . . . . . . . . . . . . . . . . . This alphanumeric data field is

Strany 83 - 3.2 Tutorials

8 - 19IPN 074-289LComposer Operating Manual8.3.5.9 Carrier GasName: . . . . . . . . . . . . . . . . . . . . . An alphanumeric data field is only

Strany 84

8 - 20 IPN 074-289L Composer Operating Manual8.3.6 Sweep ModeSweep mode is not available in Setup Tool. This mode is used primarily as a setup and d

Strany 85 - 3 - 23

8 - 21IPN 074-289LComposer Operating Manual8.3.7 Amplitude Track ModeAmplitude Track is the instrument’s default operating mode, as shown in Figure

Strany 86 - 3 - 24

8 - 22 IPN 074-289L Composer Operating Manual8.3.8 Control StateThe Control State menu choices under Action item in the menu bar allow access to con

Strany 87 - 3 - 25

8 - 23IPN 074-289LComposer Operating ManualFigure 8-8 Manual Flow Control DialogNOTE: Control, Stop, Hold, Bypass and Manual are mutually exclusive

Strany 88

8 - 24 IPN 074-289L Composer Operating Manual8.3.9 Setup Data Log FileThis feature is not available in Setup Tool. In the Amplitude Track and the Sw

Strany 89 - 3 - 27

8 - 25IPN 074-289LComposer Operating Manual8.3.10 Setup Data DisplayThis option is located in the menu item Setup >> Data Display or can be a

Strany 90 - 3 - 28

TOC - 8 IPN 074-289L Composer Operating ManualChapter 9BibliographyChapter 10GlossaryChapter 11Recommended Parts List11.1 Replacement Parts. . . . .

Strany 91 - 3 - 29

8 - 26 IPN 074-289L Composer Operating Manual8.3.11 Gas Properties DatabasePressing the button on the Toolbar or selecting from menu item Tools >

Strany 92 - 3 - 30

8 - 27IPN 074-289LComposer Operating Manual8.3.12 Gas Concentration Calculator This feature is not available in Setup Tool. The Gas Concentration

Strany 93 - 3 - 31

8 - 28 IPN 074-289L Composer Operating Manual8.3.13 I/O StatusThe I/O Status menu item allows easy readout of the status of back panel system I/O in

Strany 94 - 3 - 32

8 - 29IPN 074-289LComposer Operating Manual8.4 Errors and Warnings ReportingErrors and warnings reporting is identical in the Composer Setup Tool a

Strany 95 - changes or grounding

8 - 30 IPN 074-289L Composer Operating Manual8.5 Multiple ComposersMonitor software and Setup Tool both will support multiple Composers on one compu

Strany 96 - 3 - 34

9 - 1IPN 074-289LComposer Operating ManualChapter 9BibliographyA. Wajid, C. Gogol, C. Hurd, M. Hetzel, A. Spina, R. Lum, M. McDonald, R. J. Capik, &

Strany 97

9 - 2 IPN 074-289L Composer Operating ManualThis page is intentionally blank.

Strany 98 - 4.2 Calibration

10 - 1IPN 074-289LComposer Operating ManualChapter 10GlossaryAbsolute (Pressure)The pressure of a gas measured relative to a perfect vacuum.Algorith

Strany 99 - 4.2.3 How to Calibrate

10 - 2 IPN 074-289L Composer Operating ManualControl GainThe steady state process gain. This is normally quantified as the ratio of change of the con

Strany 100 - Composer Operating Manual

10 - 3IPN 074-289LComposer Operating ManualDiluentThe gas used for dilution of the bubbler flow to produce the desired concentration and flow.DVMDig

Strany 101 - Composer Operating Manual

1 - 1IPN 074-289LComposer Operating ManualChapter 1Getting Started1.1 IntroductionThe Composer system is designed to measure and generate signals a

Strany 102

10 - 4 IPN 074-289L Composer Operating ManualLorentzian A term that mathematically describes the shape of a resonance peak. This instrument uses a Lo

Strany 103

10 - 5IPN 074-289LComposer Operating ManualPIDProportionalIntegralDifferential; A type of control loop algorithm.Power Supply ModuleThe portion of t

Strany 104

10 - 6 IPN 074-289L Composer Operating ManualResonant ModeA standing acoustic wave. The lowest frequency mode is called the fundamental.Root Allan Va

Strany 105

10 - 7IPN 074-289LComposer Operating ManualTransducerThat portion of the instrument system containing the physical measuring apparatus; including; t

Strany 106 - 4 - 10

10 - 8 IPN 074-289L Composer Operating ManualThis page is intentionally blank.

Strany 107 - 4 - 11

11 - 1IPN 074-289LComposer Operating ManualChapter 11Recommended Parts List11.1 Replacement PartsTable 11-1 provides details about the recommended

Strany 108 - Ú below, and section

11 - 2 IPN 074-289L Composer Operating ManualThis page is intentionally blank.

Strany 109 - 4 - 13

12 - 1IPN 074-289LComposer Operating ManualChapter 12Quick Use Guide12.1 IntroductionThe Quick Use Guide provides a reference for data concerning c

Strany 110 - 4 - 14

12 - 2 IPN 074-289L Composer Operating ManualCONCENTRATIONOUTOUT INPOWEROPTIONRS232SYSTEM I/OSENSORCONTROLEARTH GROUNDCFGR1 CFGR2POWERRECEIVINGSENDIN

Strany 111 - IPN 074-289L

12 - 3IPN 074-289LComposer Operating ManualConcentration Control ConfigurationPOWER*MEMORY CHECKS*READ CONFIGURATION SWITCHES & INPUTS*RESTORE C

Strany 113 - 4 - 17

1 - 2 IPN 074-289L Composer Operating ManualFor the LED version of the Controller module, a software application called the Setup Tool allows the com

Strany 114 - 4 - 18

12 - 4 IPN 074-289L Composer Operating ManualThis page is intentionally blank.

Strany 115 - 4 - 19

Index - 1IPN 074-289LComposer Operating ManualIndexAaccuracy 3-2acousticgenerator problems 6-3noisy environment 6-4receiver problems 6-3Algorithm 10

Strany 116 - 4 - 20

Index - 2 IPN 074-289L Composer Operating ManualDD connectors 1-12damped curve 3-24data terminal 7-1DB-25 connector 4-22DC power 1-13Dead Time 3-22,

Strany 117 - 4 - 21

Index - 3IPN 074-289LComposer Operating ManualKKALREZ 1-5, 3-9LLaplace transform 2-6LCD Display 4-2, 4-9, 4-17leak detection 5-2primary containment

Strany 118 - 4 - 22

Index - 4 IPN 074-289L Composer Operating ManualRepair Support 1-4reproducibility 1-3, 4-2Resonant Chamber 1-10, 2-4, 3-6, 10-5temperature 1-11resona

Strany 119 - 4 - 23

Index - 5IPN 074-289LComposer Operating ManualWWARNING 4-8, 4-12wavelength 3-6

Strany 120 - 4.3.6.2 SENSOR Connector

Index - 6 IPN 074-289L Composer Operating ManualThis page is intentionally blank.

Strany 121 - 4 - 25

1 - 3IPN 074-289LComposer Operating Manual1.3 Performance SpecificationsThe following sections provide the Composer’s performance specifications.Tr

Strany 122 - 4 - 26

1 - 4 IPN 074-289L Composer Operating Manual1.4 How To Contact Customer SupportIf you have a question about your instrument, please read this Operat

Strany 123

1 - 5IPN 074-289LComposer Operating Manual1.4.3 Returning Your InstrumentDo not send your instrument without first speaking with a Customer Support

Strany 124 - 4 - 28

1 - 6 IPN 074-289L Composer Operating Manual1.6 Physical RequirementsThe following sections provide the physical dimensions, weight, mounting requir

Strany 125 - 4.3.7 Parameter Entry

1 - 7IPN 074-289LComposer Operating Manual1.7 Electrical Power RequirementsThe following sections detail the electrical power requirements for the

Strany 126 - 4 - 30

1 - 8 IPN 074-289L Composer Operating Manual1.7.2 Rated InputOperational Voltage Range . . . . . . 90 to 264 V(ac)Overvoltage . . . . . . . . . . .

Strany 127 - Maintenance

1 - 9IPN 074-289LComposer Operating Manual1.8 Environmental RequirementsThe following paragraphs detail the Composer’s environmental requirements.U

Strany 128

1 - 10 IPN 074-289L Composer Operating Manual1.10.1 Physical InstallationThe Transducer is connected to the Reactor’s delivery tubing with Cajon™ VC

Strany 129 - Secondary Containment Vessel

1 - 11IPN 074-289LComposer Operating Manual1.10.1.1 Use with Heated Delivery LinesThe Transducer may be attached to Delivery Lines heated to a maxi

Strany 130

OPERATING MANUALCOMPOSERGas Concentration ControllerTWO TECHNOLOGY PLACEEAST SYRACUSE, NY 13057-9714 USAPhone: +315.434.1100Fax: +315.437.3803Email:

Strany 131

1 - 12 IPN 074-289L Composer Operating Manual1.10.2.1.1 The BNC ConnectorsBecause complete BNC cables are so common, there are no mating connectors

Strany 132 - (LCD Display Option Only)

1 - 13IPN 074-289LComposer Operating ManualFigure 1-2 Solder Cup Connector1.10.2.3 How to Connect Electrical Power to the InstrumentAll external

Strany 133 - Diagnosing Problems

1 - 14 IPN 074-289L Composer Operating Manual1.10.2.5 Protective Grounding RequirementsThe Controller and Transducer must be connected to a low impe

Strany 134

1 - 15IPN 074-289LComposer Operating Manual1.10.3.2 Connections To Earth GroundThe ground connection on the Controller is a threaded stud with a he

Strany 135

1 - 16 IPN 074-289L Composer Operating Manual1.10.3.3 Minimizing Noise Pickup from External CablingWhen an instrument is fully integrated into a dep

Strany 136

1 - 17IPN 074-289LComposer Operating Manual1.10.4 Connections to the ReactorWhen integrating this Instrument System there are three basic levels of

Strany 137

1 - 18 IPN 074-289L Composer Operating Manual1.10.4.1 When Used as a MonitorConcentration output is available from an analog voltage output, see sec

Strany 138

1 - 19IPN 074-289LComposer Operating Manual1.10.4.2 Basic Connection to a Reactor (Concentration Measuring Mode)If a permanent connection from the

Strany 139 - 6.5 User Diagnostics

1 - 20 IPN 074-289L Composer Operating Manual1.10.4.3 How to Connect the Instrument for Independent Control of the ConcentrationThis Instrument Syst

Strany 140

1 - 21IPN 074-289LComposer Operating ManualThe parameters necessary for tuning the concentration maintaining Control Loop are readily visible and al

Strany 141

TrademarksThe trademarks of the products mentioned in this Operating Manual are held by the companies that produce them.INFICON® and Composer® are tra

Strany 142 - 6.6 Error Detection

1 - 22 IPN 074-289L Composer Operating Manual1.10.5.2 PositioningThere is no preferred orientation for the Transducer. It is, however, best to maint

Strany 143 - External Communications

1 - 23IPN 074-289LComposer Operating Manual1.10.5.5 Transducer SeriesComposer transducers are divided into two series. The controller must be confi

Strany 144 - 7.4 Basic Command Structure

1 - 24 IPN 074-289L Composer Operating ManualThis page is intentionally blank.

Strany 145

2 - 1IPN 074-289LComposer Operating ManualChapter 2How the Instrument Works2.1 Speed of Sound and Gas CompositionThe speed of sound, C, in an ideal

Strany 146

2 - 2 IPN 074-289L Composer Operating Manual[6]In practice, the exact value of L is unimportant. Composition is computed from knowledge of the mixtur

Strany 147

2 - 3IPN 074-289LComposer Operating ManualCombining 12, 13 and 14:[15]Simplifying leads to,[16]where, [17][18][19]Since the equation is obviously a

Strany 148

2 - 4 IPN 074-289L Composer Operating ManualFigure 2-1 Mole Fraction of Trimethylindium in H2 vs. Instrument’s Frequency2.2 Measuring the Speed of

Strany 149

2 - 5IPN 074-289LComposer Operating Manual(Manipulated Variable). It is normal for this temperature to be controlled to +/- 0.05 °C. It is also norm

Strany 150

2 - 6 IPN 074-289L Composer Operating Manual2.3 Control Loop PrimerThe instrumental advances in measurement speed, precision and reliability would n

Strany 151

2 - 7IPN 074-289LComposer Operating ManualFigure 2-2 Response of process to an open loop step change at t=0 (control signal is increased)[23][24]Th

Strany 152 - 7.4.4 Error Codes

DECLARATION OF CONFORMITYThis is to certify that this equipment, designed and manufactured by:INFICON Inc.2 Technology Place

Strany 153 - 7 - 11

2 - 8 IPN 074-289L Composer Operating Manualcontrollers. R(s) represents the concentration set point. The feedback mechanism is the error generated b

Strany 154 - 7.4.5 Examples

2 - 9IPN 074-289LComposer Operating Manual[28]ITAE is insensitive to the initial and somewhat unavoidable errors, but it will weight heavily errors

Strany 155 - Software Guide

2 - 10 IPN 074-289L Composer Operating ManualOr, If the change-in-error is negative (moving away from set point), then the action is large negative

Strany 156 - 8.2.2 Software Installation

2 - 11IPN 074-289LComposer Operating ManualFigure 2-4 Comparison of PID and Fuzzy Logic ControlPID Control:Process Gain = 0.96, Dead Time = 14 sec.

Strany 157 - 8.2.3 First Time Use

2 - 12 IPN 074-289L Composer Operating Manual2.5 Instrument OverviewThe Instrument System Block Diagram, see Figure 2-5, depicts the general layout

Strany 158

2 - 13IPN 074-289LComposer Operating Manual2.6 Description of SubsystemsThe two major components which comprise the Composer are the Transducer and

Strany 159

2 - 14 IPN 074-289L Composer Operating ManualFigure 2-6 . Transducer Block Diagram

Strany 160 - 8.3 Software Reference

2 - 15IPN 074-289LComposer Operating Manual2.6.2 ControllerIPN 761-600The following sections describe the 761-600 controller; see the block diagram

Strany 161

2 - 16 IPN 074-289L Composer Operating Manual2.6.2.3 FPGA U33The FPGA is an Actel A1010B programmed to act as an interface to the various sections o

Strany 162

2 - 17IPN 074-289LComposer Operating Manual2.6.2.5 Sensor Temperature ControlU3, U5D, U13, U14, VR1, U19, Q2Temperature of the sensor is measured b

Strany 164 - 8.3.2 Toolbar

2 - 18 IPN 074-289L Composer Operating Manual2.6.2.7 Power SupplyU27, U25, U23, U11,U10, U22, U4A LT1070(U27) is used as an 18 volt pre-regulator ad

Strany 165 - 8 - 11

2 - 19IPN 074-289LComposer Operating ManualFigure 2-7 Electrical System Block Diagram

Strany 166 - 8.3.3 Ribbon Bar

2 - 20 IPN 074-289L Composer Operating ManualThis page is intentionally blank.

Strany 167 - Table 8-4 Ribbon Bar Buttons

3 - 1IPN 074-289LComposer Operating ManualChapter 3Applications Guide3.1 Advice and TipsThe following paragraphs review many of the common question

Strany 168 - 8 - 14

3 - 2 IPN 074-289L Composer Operating ManualNOTE: Always wait for the STEADY lamp on the Controller’s front panel to illuminate before completing the

Strany 169 - 8 - 15

3 - 3IPN 074-289LComposer Operating Manualto a range of about 1.1 to 1.7 for almost every known gas. Consequently, guesses that are wrong don’t make

Strany 170 - 8 - 16

3 - 4 IPN 074-289L Composer Operating ManualNOTE: Long term use of the Composer with some of the materials listed in Table 3-1 may not be possible. T

Strany 171 - 8 - 17

3 - 5IPN 074-289LComposer Operating Manual3.1.6 What is the Effect of Pressure Variation?Because not all mixtures are purely ideal and the use of D

Strany 172 - 8 - 18

3 - 6 IPN 074-289L Composer Operating Manual3.1.8 What is the Effect of Flow Variation?There is virtually no flow induced effect on concentration me

Strany 173 - 8 - 19

3 - 7IPN 074-289LComposer Operating Manual3.1.9.1 Calculation ExampleThe time interval for an acoustic pulse to travel 0.2 meters in hydrogen is co

Strany 174 - 8.3.6 Sweep Mode

Registration CardTWO TECHNOLOGY PLACEEAST SYRACUSE, NY 13057-9714 USAPhone: +315.434.1100Fax: +315.437.3803Email: [email protected]. BOX 1000FL-9

Strany 175 - 8.3.7 Amplitude Track Mode

3 - 8 IPN 074-289L Composer Operating Manual3.1.10 What is the Long Term Stability of this Measurement SystemA Composer Transducer was checked weekl

Strany 176 - 8.3.8 Control State

3 - 9IPN 074-289LComposer Operating Manual3.1.11 Properties of Optional O-ring MaterialsThe use of some chemicals may preclude the use of Viton sea

Strany 177 - 8 - 23

3 - 10 IPN 074-289L Composer Operating Manual3.1.12 What Happens if the Reference Zero is Improperly SetIf the Reference Zero is set when the condit

Strany 178 - 8.3.9 Setup Data Log File

3 - 11IPN 074-289LComposer Operating Manual3.1.13 Composer Bulk Flow Control Configuration3.1.13.1 GeneralComposers with firmware version 4.00 (LE

Strany 179 - 8.3.10 Setup Data Display

3 - 12 IPN 074-289L Composer Operating Manual3.1.13.3 Controller Installation (Bulk Flow)SW#8 of CFGR2 must be activated to use Bulk flow Control. S

Strany 180 - 8 - 26

3 - 13IPN 074-289LComposer Operating ManualThe Analog I/O option card has a 9 pin D connector. The pin out is as follows:The System I/O connector is

Strany 181 - 8 - 27

3 - 14 IPN 074-289L Composer Operating ManualFigure 3-7 . Bulk Flow installation scheme showing optional Bulk Flow Cable’s details. 4.99 megohm resi

Strany 182 - 8.3.14 About

3 - 15IPN 074-289LComposer Operating Manual3.1.13.4 Operation (Bulk Flow)The PROG key alternates the display between operate and program modes. In

Strany 183 - 8 - 29

3 - 16 IPN 074-289L Composer Operating Manual3.1.13.5 Programming the Controller’s Parameters (Bulk Flow)When operating in the “Bulk Flow Control” c

Strany 184 - 8.5 Multiple Composers

3 - 17IPN 074-289LComposer Operating ManualMinimum Bubbler Flow (User 3) . . Parameter used to set a minimum opening for the Bubbler’s MFC, avoidi

Strany 185 - Bibliography

BUSINESS REPLY MAILFIRST CLASS PERMIT NO. 49 EAST SYRACUSE, NEW YORKPOSTAGE WILL BE PAID BY ADDRESSEEINFICON INC.Two Technology PlaceEast Syracuse, Ne

Strany 186

3 - 18 IPN 074-289L Composer Operating Manual3.1.13.6 Determining Process Control Loop Parameters (Bulk Flow)3.1.13.6.1 Gain, KpThe most important

Strany 187 - Glossary

3 - 19IPN 074-289LComposer Operating Manual3.1.14 Analog I/O Option — Concentration Control ConfigurationIn addition to its use enabling the Bulk F

Strany 188 - 10 - 2

3 - 20 IPN 074-289L Composer Operating Manual3.1.15 How to Sample Flow with a ComposerWhen the flow through a pipe exceeds the Transducer’s 1800 scc

Strany 189 - 10 - 3

3 - 21IPN 074-289LComposer Operating Manual3.2 TutorialsThe following information is available: Control Loop Fundamentals — Concentration Control,

Strany 190 - 10 - 4

3 - 22 IPN 074-289L Composer Operating ManualNOTE: The use of a chart recorder, especially when beginning a new application, is highly recommended. C

Strany 191 - 10 - 5

3 - 23IPN 074-289LComposer Operating ManualFigure 3-9 Examples of Delay, or Deadtime Settings3.2.1.2 Tuning a Fast Delivery SystemIf the source re

Strany 192 - 10 - 6

3 - 24 IPN 074-289L Composer Operating ManualThe response of a system with too little controller gain (its PROCESS GAIN value is too large) is charac

Strany 193 - 10 - 7

3 - 25IPN 074-289LComposer Operating Manual3.2.1.5 Computing the System’s Time ConstantThis Instrument System operates at 1 Hz in the Amplitude Tra

Strany 194 - 10 - 8

3 - 26 IPN 074-289L Composer Operating Manual7 Start data gathering. NOTE: Zeroing the instrument is not essential, but the concentration reading sho

Strany 195 - 11.1 Replacement Parts

3 - 27IPN 074-289LComposer Operating Manual21 Use linear regression technique to fit data to Y(t) = A + Bt. Then, by comparing with the above equati

Strany 196 - 11 - 2

WarrantyTWO TECHNOLOGY PLACEEAST SYRACUSE, NY 13057-9714 USAPhone: +315.434.1100Fax: +315.437.3803Email: [email protected]. BOX 1000FL-9496 BALZE

Strany 197 - Quick Use Guide

3 - 28 IPN 074-289L Composer Operating ManualFigure 3-12 Linear regression fit of data points taken in Figure 3-11. In this particular process, the

Strany 198 - 12 - 2

3 - 29IPN 074-289LComposer Operating Manual3.2.3.1 Using Composer Monitor Software / Setup Tool1 From the Setup >> Process Parameters menu op

Strany 199 - 12 - 3

3 - 30 IPN 074-289L Composer Operating Manual15 Enter these values through the menu Setup >> Process Parameters. Also, select the control loop

Strany 200 - 12 - 4

3 - 31IPN 074-289LComposer Operating Manual3.2.3.3 Using Composer’s Front Panel Keypad (LCD Display Option)1 Press the PROG button on the keypad to

Strany 201 - Index - 1

3 - 32 IPN 074-289L Composer Operating Manual14 Enter the programming mode again. Modify the process parameters with newly found values. Also, select

Strany 202

3 - 33IPN 074-289LComposer Operating Manual3.2.5 Why is a Digital Connection to the System Controller Preferred?Digital information is not corrupte

Strany 203 - Index - 3

3 - 34 IPN 074-289L Composer Operating ManualThis page is intentionally blank.

Strany 204 - Index - 4

4 - 1IPN 074-289LComposer Operating ManualChapter 4OperationNOTE: If you are experienced with operating this instrument, Chapter 12, Quick Use Guide

Strany 205 - Index - 5

4 - 2 IPN 074-289L Composer Operating Manual4.1.2 Optional LCD DisplayThe parameter set may be downloaded to the instrument in any of the methods li

Strany 206 - Index - 6

4 - 3IPN 074-289LComposer Operating Manual4.2.3 How to CalibrateFlow the calibration gas (usually pure carrier gas) through the Transducer at the n